Caution:
While the parts counting function is active the functionality of
button
is changed. Pressing
does not result in printing and saving
weighments.
26.2.2. Minimum reference mass
It concerns the <
Parts counting > working mode
Users before beginning the procedure of single piece mass evaluation can
declare
“minimum reference mass”
i.e. minimum total weight of all pieces
put on the scale pan expressed in reading divisions.
Procedure:
•
Enter parameter group
<
Working modes>
according to ch.
24 of this manual,
•
Choose
„
Parts counting /
Minimum reference mass”
and then set an appropriate value.
Accessible settings:
1 d, 2 d, 5 d, 10 d.
Caution:
While the procedure of evaluation the mass of single piece the mass
of all pieces put on the pan is lower than the value declared in parameter
„
Minimum reference mass”, the following warning message will
be displayed: < Too low sample mass>.
26.3. Setting a reference unit by entering known piece mass
Procedure:
•
Enter
mode
<
Parts counting>
according to ch. 26.1
of this manual,
•
Press (enter piece mass), then an editing field is displayed
<Reference Unit>
with the screen keyboard,
•
Enter a value and confirm it by pressing
, which causes starting
<
Parts counting >
with automatic setting the reference unit.
122