On carrying out the third measurement, the scale automatically
determines the density of the porous body, and indicates the result
on terminal’s display. If the user needs a report from the density
determination process, it can be printed on a plugged printer by
pressing
key. The density determination process is then
completed,
•
A report from density determination of a porous body is saved in
<
Density>
database. The report is named by process date.
31.4. Reporting from completed density determination processes
On completing each density determination process, the scale’s software
automatically generates a report.
Caution:
Submenu: “
Devices /
Printer /
Printouts /
Density
printout template” enables optional configuration of report template
(see ch. 16.2.3 of this user manual).
Default form of the report from density determination template:
--------------------------------------------
Density
--------------------------------------------
{40:Operator:,-25}{75}
{40:Start date:,-25}{155}
{40:End date:,-25}{156}
{40:Standard liquid:,-25}{158}
{40:Method:,-25}{157}
{40:Weighing 1:,-25}{165}
{40:Weighing 2:,-25}{166}
{40:Density:,-25}{162}{163}
--------------------------------------------
The report from each completed density determination process is
simultaneously saved in the database of
<
Density Reports>
, where
the files are named by their date and hour of process execution.
List of
data for the density determination process – see ch. 35.5.4 of this user
manual.
184