The user can export/import scale settings (printout templates, user
parameters) using flash memory stick.
Procedure for settings export:
•
Connect flash memory stick to USB port of your scale,
•
Enter
„
Others /
Export”
submenu, settings are exported
automatically to flash memory stick plugged to scale USB port.
•
Upon completed procedure the following message is displayed:
„Operation completed successfully”
.
Procedure for settings import:
•
Connect flash memory stick to USB port of your scale,
•
Enter
„
Others /
Import”
submenu, settings are imported
automatically from flash memory stick plugged to scale USB port.
•
Upon completed procedure the following message is displayed:
„Operation completed successfully”
.
22. USER ADJUSTMENT
An option only for non-verified scale
Scales require to recalculate internal divisions to more suitable ones (e.g.
g, kg etc.). In order to do this they require an adjustment factor. It is adjusted
during the adjustment procedure using a mass standard. Adjustment should
be carried out if weighing a standard mass shows a different mass value.
To enter
<
User Adjustment>
,
press
key and then
:
„
User Adjustment”
.