11.3. Zeroing
To
ZERO
the scale press:
.
The scale will display zero and following pictograms:
and
.
Zeroing is only possible within the scope of
±2%
of full scale. While
zeroing outside the scope of
±2%
you will see
<Err2>
. Zeroing is
possible only in stable state.
Notice:
Zeroing is possible only within the ±2% interval of the maximal range.
If zeroing is performed beyond this range the <Err2> message and
short audible signal will be emitted.
11.4. Weighings in two ranges
Switching between the
I range
and the
II range
happens automatically
(exceeding Max of the
I range
). Weighings in the second range is
signalled by a pictogram in the top left corner of the display.
Then weighings is done with the accuracy of the
II range
to
the moment
of returning to zero (autozero range
)
where the scale switches back
to the
I range
.
11.5. Selection of basic weight unit
This function is used to set weight unit the scale will start with.
Procedure:
•
Enter the submenu
<P3.Unit>
and then: