- 26 -
12.1. Setting a filtering level
Procedure:
•
Enter the submenu
<P1.rEAd>
and then:
1 - 4
- level of filtering
•
By pressing
select the filtering level you need
Notice:
Filtering level influences the time of stabilization. The higher the filtering
level is the longer stabilization time is needed.
Return to weighing:
See - 10.2.2.
12.2. Median filter
This filter eliminates short changes (impulses) of measure signal
(e.g. shocks).
Procedure:
•
Enter the submenu
<P1.rEAd>
and then: