- 31 -
Pr_n
noStAb
- immediate printout
(not accessible in verified scales)
Pr_n
StAb
- sending stable results
Pr_n
rEPL
- automatic operation
Pr_n
CntA
- continuous transmission in basic unit
Pr_n
Cntb
- continuous transmission in present unit
Return to weighing:
see 10.2.2.
13.2. Minimal mass threshold
This function is necessary while working with
automatic tare
or
automatic operation or weighing animals
.
Automatic tarring
will not be applied until the indication (gross)
is lower than the value inscribed in
S_Lo
parameter.
In automatic operation
measurements (net) are sent via RS232
when the indication is equal or greater than the value inscribed in
S_Lo
parameter.
Weighings animals
is performer when the indication is equal
or greater than the value inscribed in
S_Lo
parameter.
Procedure:
•
Enter the submenu
<P2.Prnt>
and then: