- 57 -
16. USER CALIBRATION
Only for non-verified scales
Confirmation of high accuracy of weighing requires periodical correcting
of calibration factors in the scale memory – this is adjustment of the scale.
Calibration should be performed when we start weighing or dynamic
change of temperature occurs. Before starting calibration remove
loads from the pan.
16.1. Calibration
Procedure:
•
Enter the submenu
<P6.CAL>
and then:
•
Following inscriptions will appear
•
A new start mass is adjusted during this period of time.
After that a mass of calibration weight is shown (e.g. 3 000kg).