- 63 -
Notice:
1. Each command have to be terminated in CR LF;
2. The best Policy for communication is not sending another
command until the former answer has been received.
19.3. Respond message format
After sending a request message you can receive:
XX_A CR LF
command accepted and in progress
XX_D CR LF
command completed (appears only after XX_A)
XX_I CR LF
command comprehended but cannot be executed
XX _ ^ CR LF
command comprehended but time overflow error appeared
XX _ v CR LF
command comprehended but the indication below the
XX _ OK CR LF
Command done
ES_CR LF
Command not comprehended
XX _ E CR LF
error while executing command – time limit for stable result
exceeded (limit time is a descriptive parameter of the scale)
XX
- command name
_
- substitutes spaces
19.4. Command’s description
19.4.1. Zeroing
Syntax
Z CR LF
Possible answers:
Z_A CR LF
Z_D CR LF
- command accepted and in progress
- command completed
Z_A CR LF
Z_^ CR LF
- command accepted and in progress
- command comprehended but zero range overflow appeared
Z_A CR LF
Z_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
Z_I CR LF
- command comprehended but cannot be executed
Summary of Contents for WPT/4
Page 2: ...2 AUGUST 2016...
Page 10: ...10 4 MAIN DIMENSIONS 4 1 Low profiled scales...
Page 11: ...11 4 2 Ramp scales 4 3 Stainless steel scales...
Page 12: ...12 4 4 Stainless steel scales pit version 4 5 Mild steel scales...
Page 13: ...13 4 6 Pallet scales 4 7 Beam scales...
Page 32: ...32 Return to weighing see 10 2 2 13 3 Baud rate Procedure Enter the submenu P2 Prnt and then...
Page 74: ...74...
Page 75: ...75...