- 65 -
Notice:
This protocol uses the dot character as a decimal point for separating
the decimal fraction part.
19.4.5. Send the stable result in basic unit
Syntax:
S CR LF
Possible answers:
S_A CR LF
S_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1
2-3
4
5
6
7-15
16
17
18
19
20
21
S
space
stability
space
sign
mass
space
unit
CR
LF
Example:
S CR LF
– computer command
S _ A CR LF
- command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
– command done,
mass value in basic unit is returned.
19.4.6. Send the result immediately in basic unit
Syntax:
SI CR LF
Possible answers:
SI_I CR LF
- command comprehended but cannot be executed at the moment
MASS FRAME
- mass value in basic unit is returned
Summary of Contents for WPT/4
Page 2: ...2 AUGUST 2016...
Page 10: ...10 4 MAIN DIMENSIONS 4 1 Low profiled scales...
Page 11: ...11 4 2 Ramp scales 4 3 Stainless steel scales...
Page 12: ...12 4 4 Stainless steel scales pit version 4 5 Mild steel scales...
Page 13: ...13 4 6 Pallet scales 4 7 Beam scales...
Page 32: ...32 Return to weighing see 10 2 2 13 3 Baud rate Procedure Enter the submenu P2 Prnt and then...
Page 74: ...74...
Page 75: ...75...