TROUBLESHOOTING
8-6
with the host. For optimal performance, use the
rechargeable Lithium-ion batteries (part # 029-3053-000).
Table 1.
Compatibility chart of ProRAE Remote and the
AreaRAE Steel Firmware
Firmware Versions
Software Version
2.30
2.31 2.32 3.00
1.0 OK
OK
NO NO
1.20 OK
OK
NO NO
1.30 OK
OK
OK
OK
1.40 OK
OK
OK
OK
ProRAE Remote Version 1.41 can support all firmware
versions; howeve, some newer firmware versions are
not compatible with ProRAE Remote Version 1.0 or
1.20.
3.
MODE OF OPERATION:
The AreaRAE must be in regular
operation mode. An antenna symbol should appear on the
AreaRAE LCD, indicating the host is attempting to pull
information from the unit. If any of the programming sub-
menus have been entered, such as “Calibrate Monitor”
then communication will fail.
4.
RADIO
: The “Radio” button on all sensing units must be
activated. If pushing the “Radio” button does not make the
red LED appear, then contact RAE Systems Service at
tech@raesystems.com or 888-723-4800.
5.
COM PORT:
The same communication port that the
RS232 cable is plugged into must be selected in the
software. Within the ProRAE Remote proceed to “Set Up”
(A)>> “Host” >>“RF Modem Port” (B) (Figure 2). Refer to
Technical Note TN-170 for more detailed setup instructions
including using USB ports. The chosen communication
port must be functioning properly and cannot be occupied
by another program. To verify the functionality of the
communication port, RAE Systems Com Port Checker Kit
(part number 029-0005-000) can be used. Simply follow
the instructions provided with the Checker Program.