- READ BEFORE OPERATING -
This manual must be carefully read by
ALL
individuals who
have or will have the responsibility for using, maintaining, or
servicing this product. The product will perform as designed
only if it is used, maintained, and serviced in accordance with
the manufacturer's instructions.
When the IAQRAE Monitor (PGM-5210) is removed from the
transport case and turned on for the first time, there may be
residual organic or inorganic vapor trapped inside the detector
chambers. The initial PID or toxic gas sensor reading may
indicate a few ppm. Ensure an area free of organic or toxic
vapors and turn the monitor on. After running the monitor for
several minutes, the residual vapor in the detector chamber
should clear, and the reading should return to zero.
The battery of the IAQRAE Monitor will slowly drain even when
it is not on. If the monitor has not been charged for 5-7 days,
the battery voltage will be low. Therefore, it is good practice to
charge the monitor before each use. It is also recommended to
fully charge the monitor for AT LEAST 10 HOURS before initial
use. See Section 3.1 for more information on battery charging
and replacement.
CAUTION!
To reduce the risk of electric shock, turn the power off before
removing the monitor cover. Disconnect the battery before
removing sensor module for service. Never operate the monitor
when the cover is removed. Remove monitor cover and sensor
module only in an area known to be non-hazardous.