MultiRAE
User’s Guide
54
16.
Once you have made your selections in ProRAE Studio II, you must upload the changes to the
instrument. Click the icon labeled “Upload all settings to the instrument.”
17.
A confirmation screen is shown. Click “Yes” to perform the upload, or “No” to abort.
Uploading takes a few seconds, and this progress bar is shown. You can abort the upload by clicking
“Cancel.”
18.
Exit ProRAE Studio II.
19.
Press [Y/+] on the MultiRAE to exit Communication Mode.
9.2 Deactivating Policy Enforcement
9.2.1 AutoRAE 2 Cradle
To deactivate Policy Enforcement when using an AutoRAE 2 Cradle, follow the procedure for changing
settings. See page 49 for details.
9.2.2 MultiRAE Desktop Cradle Or Travel Charger
If the MultiRAE screen displays the message that it must be bump tested or calibrated, and if the option to
bypass bump testing or calibration is not available, you should shut off the instrument and follow the
procedure outlined here if you want to change the Policy Enforcement settings:
1.
Use a USB cable to connect the MultiRAE in its Travel Charger or Desktop Cradle to a computer
running ProRAE Studio II.
2.
Enter Diagnostic Mode on the MultiRAE (with the instrument turned off, press and hold [Y/+] and
[MODE] until it starts up.
3.
After startup, enter the password when prompted (default is “0000”) and press [MODE].
4.
Press [N/-] repeatedly until you see the “Enter Communications Mode?” screen.
5.
Press [Y/+] to enter Communications Mode.
6.
Start ProRAE Studio II.
7.
Select “Administrator.”