ppbRAE 3000 User’s Guide
Alarm Signal Summary
Message Condition
Alarm Signal
HIGH
Gas exceeds “High
Alarm” limit
3 beeps/flashes per second*
OVR
Gas exceeds
measurement range
3 beeps/flashes per second*
MAX
Gas exceeds electronics’
maximum range
3 beeps/flashes per second*
LOW
Gas exceeds “Low
Alarm” limit
2 beeps/flashes per second*
TWA
Gas exceeds “TWA”
limit
1 Beep/flash per second*
STEL
Gas exceeds “STEL”
limit
1 Beep/flash per second*
Pump
icon
flashes
Pump failure
3 beeps/flashes per second
Lamp
PID lamp failure
3 beeps/flashes per second
plus “Lamp” message on
display
Battery
icon
flashes
Low battery
1 flash, 1 beep per minute
plus battery icon flashes on
display
CAL
Calibration failed, or
needs calibration
1 beep/flash per second
NEG
Gas reading measures
less than number stored in
calibration
1 beep/flash per second
* Hygiene mode only. In Search mode, the number of beeps per
second (1 to 7) depends upon the concentration of the sampled gas.
Faster rates indicate higher concentrations.
27
Summary of Contents for ppbRAE 3000
Page 1: ...ppbRAE 3000 User s Guide Rev C August 2010 P N 059 4021 000 ...
Page 27: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 25 ...
Page 45: ...ppbRAE 3000 User s Guide 43 Note Dashed line indicates automatic progression ...
Page 66: ...ppbRAE 3000 User s Guide 64 Note Dashed line indicates automatic progression ...
Page 83: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 81 ...
Page 85: ...ppbRAE 3000 User s Guide 83 Note Dashed line indicates automatic progression ...
Page 114: ......