ppbRAE 3000 User’s Guide
Zero (Fresh Air) Calibration
This procedure determines the zero point of the sensor calibration
curve. To perform a fresh air calibration, use the calibration adapter to
connect the instrument to a “fresh” air source such as from a cylinder
or Tedlar bag (optional accessory). The “fresh” air is clean, dry air
without organic impurities and an oxygen value of 20.9%. If such an
air cylinder is not available, any clean ambient air without detectable
contaminants or a charcoal filter can be used.
At the Zero Calibration menu, you can proceed to perform a Zero
calibration or bypass Zero calibration and perform a Span calibration.
You may also go back to the initial Calibration menu if you want to
exit calibration.
•
Press [Y/+] to start calibration.
•
Press [MODE] to quit and return to the main calibration
display.
If you have pressed [Y/+] to enter Zero calibration, then you will see
this message:
1.
Turn on your Zero calibration gas.
2.
Press [Y/+] to start calibration.
Note:
At this point, you may press [MODE] if you decide
that you do not want to initiate calibration. This will take you
directly to the Calibration menu, highlighted for Span
calibration.
37
Summary of Contents for ppbRAE 3000
Page 1: ...ppbRAE 3000 User s Guide Rev C August 2010 P N 059 4021 000 ...
Page 27: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 25 ...
Page 45: ...ppbRAE 3000 User s Guide 43 Note Dashed line indicates automatic progression ...
Page 66: ...ppbRAE 3000 User s Guide 64 Note Dashed line indicates automatic progression ...
Page 83: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 81 ...
Page 85: ...ppbRAE 3000 User s Guide 83 Note Dashed line indicates automatic progression ...
Page 114: ......