ppbRAE 3000 User’s Guide
Diagnostic Mode
IMPORTANT!
Diagnostic Mode is designed for servicing and
manufacturing, and therefore is not intended for everyday use, even
by advanced users. It provides raw data from sensors and about
settings, but only allows adjustment of pump stall parameters, which
should only be changed by qualified personnel.
Note:
If the instrument is turned on in Diagnostic Mode and you
switch to User Mode, datalog data remains in raw count form. To
change to standard readings, you must restart the instrument.
Entering Diagnostic Mode
Note:
To enter Diagnostic Mode, you must begin with the instrument
turned off.
Press and hold [Y/+] and [MODE] until the instrument starts.
The instrument goes through a brief startup, and then displays raw
data for the PID sensor. These numbers are raw sensor readings
without calibration. The instrument is now in Diagnostic Mode.
Note:
In Diagnostic Mode, the pump and lamp are normally on.
You can enter Programming Mode and calibrate the instrument as
usual by pressing both [MODE] and [N/-] for three seconds.
You can enter Monitoring Mode by pressing [MODE] and [Y/+]
together for three seconds.
Once the instrument is started up in Diagnostic Mode, you can switch
between Diagnostic Mode and Monitoring Mode by pressing and
holding [MODE] and [Y/+] simultaneously for two seconds.
In Diagnostic mode, you can step through parameter screens by
pressing [MODE].
85
Summary of Contents for ppbRAE 3000
Page 1: ...ppbRAE 3000 User s Guide Rev C August 2010 P N 059 4021 000 ...
Page 27: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 25 ...
Page 45: ...ppbRAE 3000 User s Guide 43 Note Dashed line indicates automatic progression ...
Page 66: ...ppbRAE 3000 User s Guide 64 Note Dashed line indicates automatic progression ...
Page 83: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 81 ...
Page 85: ...ppbRAE 3000 User s Guide 83 Note Dashed line indicates automatic progression ...
Page 114: ......