background image

QRAE 3 

Diffusion & Pumped 

QuickStart Guide 

 

 
 

 

 

 

 

 
 

 
 
 
 
 

 

 

WARNINGS 

Read Before Operating

 

The QRAE 3 User’s Guide must be carefully read by all 
individuals who have or will have the responsibility of 
using, maintaining, or servicing this product. The product 
will perform as designed only if it is used, maintained, and 
serviced in accordance with the manufacturer’s instructions. 

 

CAUTION! 

Never operate the monitor when the rear cover is re-
moved. Remove rear cover, sensors, and/or battery only 
in an area known to be non-hazardous. Never use the 
instrument with the calibration adapter installed, as this 
can cause distorted readings, a potential safety threat. 

 

 

User Interface 

 

The QRAE 3’s user interface consists of the 
display and two keys, [MODE] and [Y/+]. The 
flippable LCD displays information such as monitored 
threats,  real-time readings and measurement units, 
alarm type (when in alarm, including cal. overdue), 
battery status, datalog (if on), and radio and connection 
quality (if available). 

 

 
 
 

 

Charging The QRAE 3 (non-ATEX/IECEx) 

 

Always fully charge the battery before use. Contacts on 
the bottom of the QRAE 3 meet the Travel Charger’s 
contact pins, transferring power. Make sure the charger 
and QRAE 3 are firmly attached. Then connect the AC 
Adapter’s barrel plug to the charger, and plug its 
transformer into an AC outlet. While charging, the 
LED on the Travel Charger glows red. When the 
battery is fully charged, the LED glows green. 

 

 

 

 

 
 
 
 

Barrier Box For ATEX/IECEx Use 

 

For ATEX/IECEx applications, plug the cord from 
the Barrier Box (P/N M02-3011-000) into the 
Travel Charger, and then plug the power adapter 
into the Barrier Box. 
 
 
 
 
 
 

 

Turning The QRAE 3 On 

 

With the instrument turned off, press and hold the 
[MODE] key until the audible alarm stops, and then 
release. During startup, the battery, buzzer, vibra-
tion alarm, and LEDs are tested, and then the 
QRAE 3 performs self-testing of its other func-
tions. When the main measurement screen appears, 
the QRAE 3 is ready for calibration or use. 

 
Note: 

If the battery is completely empty, then the 

display briefly shows the message “Battery 
Discharged!” and the QRAE 3 shuts off. You should 
charge or replace the battery before turning it on again. 
 

Note:

 If “Zero At Start,” “Must Calibrate” or 

“Must Bump” is set, you will be prompted to 
perform the appropriate calibration before using the 
instrument. 

 

2. 

3. 

5. 

6. 

4. 

Charging 
status LED 

Phone: 

408-952-8200      

Fax: 

408-952-8480 

Email:

 RAE-tech@honeywell.com 

www.raesystems.com 

RAE Systems by Honeywell 
3775 N. First St. 
San Jose, CA 95134-1708 USA 

AC adapter 
barrel plug 

QRAE 3 

Travel 
Charger 

Clip Travel Charger 
Onto QRAE 3 

Non-ATEX/IECEx: 
Power supply 
plugged directly into 
Travel Charger. 

ATEX/IECEx: 
Barrier Box between 
power supply and 
Travel Charger. 

Reviews: