Technical Note TN-204    

04/17/vk

  

 

 

 

1  

Honeywell Internal

 

RAE Systems by Honeywell    877-723-2878     raesystems.com 

 

MODES & APPLICATIONS FOR THE MULTIRAE BENZENE

INTRODUCTION

  

Benzene is a very well-known human carcinogen that is widely 

distributed throughout nature (coal, crude oil, gas and gas con-

densate). In most fuel products (gasoline, diesel, jet fuel), it ex-

ists in concentrations of up to several percent, by volume (vol%). 

Benzene is a very important industrial 

chemical, as well. It belongs to the BTEX 

aromatic family and is a liquid with a ra-

ther high vapor pressure. Due to its haz-

ardous properties, Cal/OSHA set its PEL 

(Personal Exposure Limit) as 1 ppm of 

TWA and 5 ppm of STEL (Short Term 

Exposure Limit).

1

  ACGIH recommends 

even more strict limits of 0.5 and 2.5 

ppm, respectively.

2

  There are several 

portable products on the market (from 

RAE Systems, IonScience, Dräger) that 

are able to measure such low concentra-

tions of benzene, but only the MultiRAE 

Benzene combines the ability to make 

quick benzene-specific assessment 

(“snapshot”) measurements with the ad-

vantages of a versatile multi-gas monitor (up to six gas sensors) 

with real-time wireless capability. It can support more than 20 

intelligent interchangeable sensor options. 

 

Note: This Technical Note applies only to the MultiRAE Ben-

zene monitor. For information regarding the UltraRAE 3000, re-

fer to RAE Systems Technical Note TN-127: “Benzene-Specific 

Measurements In Petroleum Hydrocarbons Using The RAE 

Systems UltraRAE 3000.” 

 

INSTRUMENT AND CARTRIDGE  

The MultiRAE Benzene is designed specifically for use with the 

replaceable benzene-specific RAE-Sep Tube Cartridge, which 

comprises six individual RAE-Sep tubes. 

 

 

 

 

 

 

The MultiRAE Benzene has two modes: TVOC (Total Volatile 

Organic Compounds) mode and Benzene mode. When used in 

TVOC mode, its display shows all sensor readings continuously. 

 

 

TVOC Mode 

Benzene Mode 

When the instrument is in benzene mode, the display prompts for 

snapshot sampling. (Refer to the MultiRAE User’s Guide for details 

on managing, calibrating, and measuring in both TVOC and Ben-

zene modes

3

.)  To perform snapshot compound-specific (ben-

zene) measurement in addition to general VOC measurement 

requires a RAE-Sep Tube Cartridge (available in packs of six: 

P/N M01-0312-000).  

 

Note: Each tube should be used for one measurement only.  

 

 

 

The RAE-Sep Tube Cartridge is designed to provide six one-

time specific measurements of benzene. This cartridge has 

unique properties and performance features not found in other 

monitors’ separation tube configurations.  It has been designed 

to provide accurate, benzene-only measurement and has only 

been evaluated and tested on the MultiRAE Benzene monitor.  

Note:

 

Failure to use the MultiRAE Benzene monitor with the 

RAE-Sep Tube Cartridge will result in inaccurate detection or 

measurement, which may include the failure to detect benzene 

and possible serious injury or death.  
For proper cartridge operation and replacement, refer to the doc-

ument “RAE-Sep Tube Cartridge For MultiRAE Benzene.”

4

 

MultiRAE Benzene 

with RAE-Sep Tube 

Cartridge 

RAE-Sep Cartridge 