ToxiRAE Pro User’s Guide
4
WARNINGS
Use only RAE Systems lithium-ion battery part number G02-3004-000 (3.7V,
1,800mAh). This instrument has not been tested in an explosive gas/air atmosphere
having an oxygen concentration greater than 21%. Substitution of components may
impair suitability for intrinsic safety. Replace batteries only in non-hazardous locations.
STATIC HAZARD:
Clean only with a damp cloth to avoid damage from electrostatic
discharge.
For safety reasons this equipment must be operated and serviced by qualified personnel
only. Read and understand this instruction manual completely before operating or
servicing.
The monitor should be calibrated every time it does not pass a bump test, but no less
frequently than every six months, depending on use and exposure to gas and
contamination, and its operational mode.
The ToxiRAE Pro monitor is factory-preconfigured to a 30-day calibration interval, so it
will go into calibration alarm if it is not calibrated for 30 days or more.
• Calibration intervals and bump test procedures may vary due to national legislation.
• RAE Systems recommends using RAE calibration gas cylinders containing the gas that
is appropriate to the sensor you are using, and in the correct concentration.