PROGRAMMING
4 - 12
4.5.2 Change Low Alarm Limit
1. The second sub-menu item allows the user to change the
Low Alarm limit. Press
[Y/+]
key. The monitor will
identify the inserted tube and display the tube name. If the
tube cannot be identified, the tube can be manually selected
by responding with
[N/-]
until the correct tube name is
displayed. Next press the
[Y/+]
key and the display will
show a flashing cursor on the left-most digit of the
previously stored Low alarm limit.
2. To modify this limit, starting from the left-most digit, use
[Y/+]
or
[N/-]
key to change the digit value and press
[MODE]
key momentarily to advance to the next digit. The
flashing digit will move on to the next digit to its right.
Press and hold
[MODE]
key for 1 second to exit data entry
mode. If there is any change to the existing value, the
display shows “Save?” Press
[Y/+]
key to accept the new
value and move to next sub-menu. Press
[N/-]
key to
discard the changes.
3.
To preserve the previously stored Low Alarm limit, press
and hold
[MODE]
key for 1 second and the monitor will
exit data entry mode and move to the next sub-menu.