PROGRAMMING OF VRAE
4 - 38
4.8 Change Sensor Configuration
In the programming mode, the users may change several
sensor related configurations of the
VRAE
monitor. Table
4.7 shows the sub-menus for setting various configurations
of the sensors.
Table 4.7
Sensor Configuration Sub-Menu
Change LEL/VOL Sensor Type? *
Enable / Disable Sensors?
Change Dilution Ratio?
Change LEL/VOL Gas Selection?*
* Note: This menu is applicable only if a combination
LEL/VOL sensor is installed.
Before these sub-menus are described, the term “Correction
Factor” (CF) needs to be explained:
Correction Factor
The LEL sensor used in VRAE is a broad band sensor (i.e., it
responds to a broad range of gases). Typically, these
sensors show a different sensitivity to different gases. The
correction factor for a specific gas is defined as:
sensitivity to a calibration gas
Correction Factor =
________________________
sensitivity to a specific gas