22
6. When test strips are exposed to extreme environmental conditions.
7. If you drop the meter.
8. After changing the batteries.
Important Control Solution Informatio
n
1. Use only KNOW SUGAR control solution.
2. Check the expiration date on the control solution bottle. Do not use if expired.
3. Control solution, meter, and test strips should come to room temperature (68-86 ºF/20-30ºC)
before testing.
4. Shake the bottle before use, discard the first drop of control solution after squeezing, wipes off
the dispenser tip to avoid contaminations. These steps ensure you will get a good sample and
an accurate result.
5. Store the control solution closed at temperatures below 30ºC (86ºF). Do not refrigerate.
NOTE :
1. There are two levels of control solution (normal and high) available to choose from. The
control solution supplied with the meter kit is normal level. Please contact with your local
distributor for other level of control solution.
2. The control solution range printed on the test strip vial is for OKmeter Control Solution only.
It is used to test meter and strip performance. It is not recommended range for your blood
glucose level.
!