24
NOTE :
1. Be sure that you are in Control Solution Mode so that the test result will not be
stored in the meter memory.
2. To avoid contaminating the control solution with the content of the test strip,
DO NOT APPLY THE CONTROL SOLUTION DIRECTLY TO THE TEST STRIP!
STEP 4.
Squeeze a drop of Control Solution Shake control solution bottle
well. Remove the cap. Squeeze the bottle, discard the first drop
and wipe off the dispenser tip with a clean tissue paper or cotton.
Squeeze the bottle again to get the second drop onto a clean
non-absorbent surface or on your fingertip first.
STEP 5.
Apply Control Solution Apply the drop to the opening of the
strip absorbent channel until the confirmation window is filled.
The meter begins to count down.
STEP 6.
Check if the Test Result is in Range After the meter counts
down form 6 to 1, the test result shows up. Compare the test
result with the range printed on the test strip vial. The result
should fall within the printed range.