RainDrop Sense Operator’s Manual
LCN 50-04343 Rev. D
7-14
I am not receiving emails, even
though I have entered an email
address.
You have entered an
invalid email address. The
system does not detect
undelivered emails.
There may be a
connectivity issue with
your network.
If you are not receiving
email messages at the end
of the run, check the email
addresses you have
entered for accuracy.
Check with your IT
department to be sure you
are properly connected.
When setting up a run,
ensure that the Send Run
Emails check box is
selected.
Click the
Send Test Email
button to see if you have
solved the problem.
I am not receiving a text
message at the end of a run
even though I have entered my
phone number into the email
field.
The RainDrop Sense
instrument does not
support text messages.
You must use a valid email
address for email
notification.
I tried to fill the oil reservoir
before the oil light came on.
It is possible to overfill the
oil reservoir and to spill oil
into internal areas of the
instrument. You may not
notice it, even when you
have done this.
For this reason,
do not
refill the oil reservoir until
the oil light comes on.
There is dirt, debris, or oil
residue on the instrument.
The instrument has not
been properly cleaned.
Clean the instrument
according to the
instructions in
.
Table 7-1: Troubleshooting the RainDrop Sense Instrument