RainDrop Sense Operator’s Manual
LCN 50-04344 Rev. D
3-2
Space Requirements
The RainDrop
®
Sense instrument is designed to be placed on a standard laboratory bench.
The Instrument Controller (IC) Workstation can be located on either the left or right side of the
instrument.
Table 3-1: RainDrop Sense Dimensions (Including the IC Workstation)
Minimum Height (door open)
28 inches (71.1 cm)
Minimum Depth
19 inches (48.3 cm)
Bench Loading Weight
73 lb (26.4 kg)
Minimum Width
29 inches (74 cm)
Note:
Refer to the Pre-Site Guide for any additional installation details.
Caution:
The RainDrop Sense instrument is sensitive to vibration and should be
placed on a bench top that is free from vibration.
Gas Connectivity
Gas source should be located in proximity to the instrument because the gas supply
connection is located on the back of the instrument. Use 4mm OD polyurethane tubing with
an in-line shutoff valve. See
for information on suppliers.