RainDrop Sense Operator’s Manual
LCN 50-04344 Rev. D
3-8
Starting Up the RainDrop Sense Instrument
Use the following startup sequence for the RainDrop Sense instrument:
1.
With the RainDrop Sense and IC Workstation turned off, ensure that all the cables
have been connected properly as described in
“Connecting Cables to the RainDrop
“Connecting Cables to the Instrument Controller”
2.
Turn on the IC Workstation (on the top of the workstation).
3.
Log on to the Windows 7 operating system using the RDT account.
Note:
Logging on using any other account results in ICS failures.
User name = RDT
No password
4.
Turn on the RainDrop Sense instrument (located on the back of the instrument).
5.
Wait 1 minute for the RainDrop Sense instrument to perform internal initialization. As
the instrument initializes, the lights on the panel illuminate to indicate its status (
see
Figure 3-4: RainDance LED Lights
6.
Launch the Instrument Control Software (ICS) application as described in
lists the LED lights that indicate the status of the instrument.
Table 3-3: LED Lights Within the RainDance Logo
LED Name and Illumination Pattern
Meaning