RainDrop Sense Operator’s Manual
LCN 50-04343 Rev. D
7-3
Keep cleaning fluids away from the objective lens.
Caution:
Use caution while cleaning the chip dock, specifically in the area of the
lenses of each instrument.
DO NOT submerge the system in liquid or pour or spray liquid onto the system.
Clean the IC workstation monitor window as described in its manufacturer's
documentation.
All technical maintenance or service must be performed by RainDance Support
personnel.
Refilling the Oil Reservoir
The oil reservoir on the RainDrop Sense instrument must be refilled from time to time. You
will know that it is time to refill the reservoir when the oil light on the front of the instrument
glows red.
Caution:
Never refill an oil reservoir unless the red light is illuminated. Failure to
comply can result in an internal overflow and potential damage to the instrument.
Figure 7-1: RainDrop Sense Oil Reservoir Fill Ports
Carrier oil low
indicator light
Carrier oil reservoir
fill port
Drive oil reservoir
fill port
Drive oil low