16
User
Selections.
Note
the
diagram
for
petMAP
g
II
user
selections.
The
available
user
selections
are
light
blue
with
black
text.
Press
and
release
to
set
the
BP
measurement
frequency
which,
when
set,
the
unit
will
sound
a
short
“beep”
and
display
a
screen
message
reminding
the
user
that
it
is
time
to
take
another
BP
measurement.
The
available
BPcycle
options
are
STAT
(BP
cycles
rapidly
for
three
minutes
then
reverts
automatically
to
a
2
min
cycle
time),
1
Min,
2
Min,
3
Min,
4
Min,
5
Min,
10
Min,
15
Min,
20
Min,
25
Min
and
30
Min.
Once
the
desired
reminder
frequency
is
selected,
the
user
must
press
the
BP
Go/Stop
button
to
begin
the
measurement
–
subsequent
measurement
reminders
will
be
displayed
at
the
selected
interval,
but
BP
determinations
are
not
started
automatically
and
must
be
initiated
by
the
user.
Press
and
hold
the
BP
cycle
button
to
turn
the
cycle
mode
OFF.
Blood
Pressure
Measurement
&
Related
Commands
Press
and
release
to
set
the
low
limit
alarm
for
systolic
pressure.
The
available
settings
are
60,
80,
100,
120,
140,
160
and
OFF.
Press
and
hold
to
turn
the
alarm
OFF.
Press
and
release
to
set
the
high
limit
alarm
for
systolic
pressure.
The
available
settings
are
150,
175,
200,
225,
250
and
OFF.
Press
and
hold
to
turn
the
alarm
OFF.
Press
and
release
to
set
the
species
and
cuff
site
which
is
displayed
at
the
top.
The
seven
available
settings
are
DOG
(ForeArm,
HindFoot
and
Tail),
CAT
(ForeArm,
HindFoot
and
Tail)
and
Optimize…None.
The
non
‐
optimized
mode
should
be
used
on
any
species
other
than
dog
or
cat
or
when
comparing
petMAP
to
Doppler
or
other
oscillometric
BP
devices.
Press
and
release
to
initiate
a
new
reading,
or
to
stop
a
BP
reading
that
is
in
progress.
If
an
SD
card
is
in
use,
press
to
capture
an
image
of
the
screen
in
a
file
on
the
SD
card.
Similarly,
the
unit
will
prompt
the
user
at
power
OFF
for
the
same
feature.
(See
Operator’s
Manual
for
the
function
of
the
small
button,
above
left.)