Physiolaser
olympic 500
Version 1.4 June 2007
6
Description of the Device / Proper Usage
Technical data
Laser therapy device for mobile and local use
Physiolaser olympic 500
Laser class
3B
Protection class
IP 21 (drop proof)
Absolute failure of the measurement of the laser
power
5%
Modulation frequency
0,1 Hz - 18 688 Hz (± 0,1%)
Precision of the electrical skin resistance
measurement
± 10 %
Risk class according to RL 93/42/EWG
IIa
Current supply Charger
Model FRIWO FW7555M/08
Weight 2
kg
Dimensions (L x B x H)
25 x 10 x 15 cm
Operating conditions
Room temperature
+ 10°C to +40°C
Relative humidity
30% to 75%
Atmospheric pressure
700 hPa to 1060 hPa
Transport and storage
Room temperature
-20° C to +40 C
Relative humidity
30% to 75%
Atmospheric pressure
700 hPa to 1060 hPa
Content of delivery
incl. Current supply charger, safety goggles, manual, therapy guide, carrying case,
”Laser Radiation” warning sign (DIN IEC 76 (CO) 6).
For the usage an optional laser probe is required :
- Laser probe
- Multi-cluster probe
- Diagnostic probe incl. Hand electrode