Instrument Tour
R&S
®
Spectrum Rider FPH
25
Getting Started 1321.0996.02 ─ 09
.............................................................................28
...................................................................................28
4.2.1
RF Input
The RF input 50
Ω
is located on the top of the R&S Spectrum Rider.
Depending on the instrument models, different RF connector is used.
For model 02, model 06 and model 13, N type connector is used. For model 26,
RPC 3.5 mm connector is used.
Connect a cable or DUT to the RF input. Use a cable to connect the DUT to the
R&S Spectrum Rider, if necessary.
Make sure not to overload the R&S Spectrum Rider when a DUT is connected.
The maximum power that is permissible at the RF input is 20 dBm (or 100 mW).
The RF input is protected from static discharges and voltage pulses by a limiting
circuit.
RF power overload
The R&S Spectrum Rider maybe loaded with up to 30 dBm (or 1 W ) for up
to three minutes. If you apply 1 W for a longer period, the R&S Spectrum
Rider may be destroyed.
Connectors of the R&S Spectrum Rider