Instrument Tour
R&S
®
Spectrum Rider FPH
67
Getting Started 1321.0996.02 ─ 09
4.7.10
Resetting the R&S Spectrum Rider
You can either preset the R&S Spectrum Rider or reset it to factory settings.
Presetting the R&S Spectrum Rider
The [PRESET] key resets the R&S Spectrum Rider to the default setup of the cur-
rently active operating mode.
This default setup allows you to define the instrument with a new configuration
based on a defined measurement parameters without using parameters from a
previous measurement unintentionally still being active.
► Press the [PRESET]
key.
Resetting the R&S Spectrum Rider
A "Reset to Factory Settings" resets the R&S Spectrum Rider to the factory
defaults.
During a reset, the R&S Spectrum Rider restores the original configuration. It also
deletes all customized datasets (limit lines, standards, channel tables, transducer
tables etc.). Instead, it will reinstall all the datasets that have been available after
delivery.
Risk of data loss
All datasets you have saved are deleted during a factory reset.
1. In the "Instrument Setup" dialog box, select the "Reset to Factory Settings"
item.
2. Confirm the entry with the rotary knob.
Configuring the R&S Spectrum Rider