Preparing for Use
R&S
®
QPS201
15
Getting Started 1179.3438.02 ─ 02
a) Establish a power supply connection between touchscreen and the primary
panel.
b) Establish a USB connection between touchscreen and primary panel.
c) Establish a DisplayPort connection between touchscreen and primary panel.
Connecting resolution station monitors
► Optional: Connect the resolution station monitors to the primary or secondary panel
(up to four).
a) Establish a LAN connection between resolution station monitor and primary
panel.
Connecting accessories
1. Optional: Connect the metal detector to the primary or secondary panel.
a) Establish a LAN connection between metal detector and primary panel.
2. Optional: Connect devices like the traffic lights to manage the queue of people to
be scanned.
If you are using one of the queue management accessories, you have to connect them
accordingly.
●
Traffic lights via one of the USB connectors.
●
Light barriers via USB (for details see the installation manual).
Connecting the components to the power supply
► Connect all devices to the mains supply with the power cord that is part of the
delivery.
For resolution stations, you have several methods to supply them with power.
● Connect them to the AC mains supply.
● Connect them to an external PoE injector (R&S
QPS-Z113).
If you are using this method, connect all resolutions stations via LAN to the
PoE injector, and the PoE injector to the mains supply.
Establishing a ground connection
1.
WARNING!
Risk of electric shock. If the voltage or frequency are greater than
230
V or 50
Hz, the leakage current of the R&S
QPS can be above the limits and
can cause personal injuries from an electric shock.
In that case, a protective ground connection must be established for each panel
before connecting them to the mains supply.
● 200 V to 230 V AC and 50 Hz
No additional protective ground needed.
● 230 V to 240 V AC or a frequency greater than 50 Hz
Additional protective ground needed.
Exception: The following two 60 Hz AC distribution systems are tested and
work without an additional ground.