Preparing for use
R&S
®
TSMA6B
25
User Manual 4901.4026.02 ─ 01
5.3.3
Connecting the R&S TSMA6B-BP
Alternatively, it is possible to power the R&S TSMA6B via the R&S TSMA6B-BP.
When the R&S TSMA6B is switched off, remove the batteries from R&S TSMA6B-BP
to prevent discharge.
Risk of instrument damage
If the R&S TSMA6B is connected with a R&S TSMA6B-BP, DC power must be connec-
ted to the R&S TSMA6B-BP.
Do not connect the DC power to the R&S TSMA6B.
Figure 5-3: DC IN connectors
1 = DC IN connector R&S TSMA6B-BP
2 = DC IN connector R&S TSMA6B
●
Switch of the R&S TSMA6B.
●
Disconnect an existing connection to the DC IN connector of the R&S TSMA6B
●
Connect R&S TSMA6B-BP and R&S TSMA6B.
●
Connect DC power supply to the DC IN connector of the R&S TSMA6B-BP.
●
Switch on the R&S TSMA6B.
To use the R&S TSMA6B with the R&S TSMA6B-BP, proceed as follows.
1. Insert the batteries into the R&S TSMA6B-BP.
Connecting power supply