Managing Scanner Device with R&S TsmeDeviceManger
R&S
®
TSMA6B
154
User Manual 4901.4026.02 ─ 01
You can update the basic FPGA or the correction data on your R&S TSMA6B directly
from the "R&S TSME Device Manager", in the "Update" tab. The currently installed ver-
sions and the newest supported version (of the FPGA) or the minimum recommended
version (of the correction data) are indicated.
Figure C-3: Tab "Updates"
To perform an update
1. In the "R&S TSME Device Manager", select the serial number of the R&S TSMA6B
to be updated.
2. In the "Updates" tab, select the "Update" button for either the "Basic FPGA" or the
"Correction Data".
When updating the correction data, you are asked to select a directory to store a
backup of your current correction data.
C.4
Aligning R&S TSMA6B Manually - "Self Alignment"
During self-alignment, the R&S TSMA6B determines alignment values for the I/Q filter,
ADC offset and the I/Q imbalance.
Aligning R&S TSMA6B Manually - "Self Alignment"