EM510
SENSe Subsystem
:FUNCtion
4065.7763.32-01.00
5.80
.
.
[:ON] <sensor_function> {,<sensor_function>}
Switch on of one or several sensor functions.
Error message:
If CONCurrent = OFF, an error
-108, "Parameter not allowed"
will be generated for ten or
more parameters.
Parameters:
<sensor_function>
is one of the following strings:
"VOLTage:AC"
switch on level measurement
"FREQuency:OFFSet"
switch on offset measurement
Note:
The offset measurement works only with signals whose signal energy may vary but must not disappear
completely. At QAM or 100 % AM-modulated signals, the offset measurement cannot provide correct
values.
Note:
With SW option EM510IM (ITU Measurement) installed, apart from level and offset the measurement
functions AM modulation index , FM frequency deviation, PM phase deviation and band width
measurement are also available.
"AM"
switch on AM modulation index measurement
"AM:POSitive"
switch on AM positive modulation index measurement
"AM:NEGative"
switch on AM negative modulation index measurement
"FM"
switch on FM frequency deviation measurement
"FM:POSitive"
switch on FM positive frequency deviation measurement
"FM:NEGative"
switch on FM negative frequency deviation measurement
"PM"
switch on PM phase deviation measurement
"BANDwidth"
switch on bandwidth measurement
*RST state:
"VOLTage:AC"
Example:
FUNCtion "VOLT:AC", "FREQ:OFFS"
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...