EM510
SENSe Subsystem
:GCONtrol
4065.7763.32-01.00
5.82
.
:GCONtrol
.
.
[:FIXed|MGC] <numeric_value>|UP|DOWN|MINimum|MAXimum
Setting the MGC value.
Parameters:
<numeric_value>
gain control factor in dBµV
-30 dBµV = no gain control -> maximum sensitivity
110 dBµV = maximum gain control -> minimum sensitivity
UP|DOWN
increase|decrease of the MGC value by the value set in the
command
SENSe:GCONtrol[:FIXed|MGC]:STEP[:INCRement]
.
MINimum|MAXimum
sets the smallest|largest MGC value
*RST state:
50 dBuV
Example:
GCONtrol 50
.
.
[:FIXed|MGC]? [MINimum|MAXimum]
Query about the MGC value.
Parameters:
none
query about the current MGC value
MINimum|MAXimum
query about the smallest|largest MGC value
Result:
Gain control
Example:
GCONtrol? -> 50
.
.
:AUTO:TIME SLOW | DEFault | FAST
.
.
:AGC:TIME SLOW | DEFault | FAST
Setting the gain control time.
Parameters:
SLOW
slow gain control time
DEFault
default gain control time
FAST
fast gain control time
*RST state:
DEF
Example:
GCONtrol:AUTO:TIME FAST
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...