Preparation for Use
EM510
4.5
4065.7763.32-01.00
4.4.3.3 Update via RS232
Switch off the EM510 before running the update program.
Start update program UPDATE.EXE under DOS and Win3.1, Win95, Win98
Select configuration file EM510P1.CFG through:
File - Open Config File
Configure the connected COM port through:
File - COM port
Start update with:
Actions - Update device
Start update program UPD32.exe under WinNT, Win2000, WinXP
Select configuration file EM510P1.CFG through:
File - Config File
Configure the connected COM port through:
Config - COM Port
Start update with:
Update - via COM
Start update process
Switch on the EM510 within 30 seconds after the update has been started on the PC.
4.4.3.4 Update via Ethernet (LAN)
In this mode you can update one or more targets using the same update software.
Switch off the target EM510 unit(s) before running the update program.
Start update program UPD32.exe
Select configuration file EM510P1.CFG through menu “File - Config File”
If your computer contains more than one network adapter, UPD32 will take the first as the default
adapter to be used for the update. To change the network adapter, select ”Network Adapter” in the
”Config” menu.
Then select ”via Ethernet” within the ”Update” menu.
Once the menu item has been selected, the system will start sending broadcast telegrams.
Switch on or reset the target EM510 unit(s)
The targets will receive the broadcast telegrams and respond with a hardware identification
telegram. If this ID corresponds to the selected configuration file, the target will be listed in the next
line.
Now you can select the desired targets from the list.
Start the update procedure with the “START” button. The target units will now be updated one by
one.
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...