93
E-02
When zeroing is over, the R&S FSH outputs the
message “Power Sensor Zero OK” and switches
back to the softkey menu for the power sensor.
Now connect the R&S FSH-Z14 or R&S FSH-Z44
between the source and the load.
The R&S FSH displays the measured forward
power level in dBm and the VSWR of the load.
To achieve maximum measurement accuracy, enter
the frequency of the signal under test.
Press the FREQ softkey.
The R&S FSH opens the entry box for the frequency.
Enter the desired frequency using the numeric
keys.
Terminate the input with the ENTER key or by
pressing the FREQ softkey again.
The R&S FSH transfers the new frequency to the
power sensor which then corrects the measured
power readings.