R&S FSMR
Measurement Receiver Mode
1166.3363.12
4.4-16
E-2
from Rohde & Schwarz, read in the calibration factor from a memory in the
sensor and automatically take this factor into account. With other devices (e.g.
HP 436 A), you have to set the calibration factor manually on the device;
however, the factor applies only to a specific frequency. In this case, you can
enter
a table of the calibration factor above the frequency via the
SENSOR CAL
FACTOR
menu; when measurements are performed, this factor is taken into
account. You then have to set the calibration factor of 100% on the power
meter.
You can enter the calibration factor for two different sensors on up to 20
frequency points each. Depending on the setting of the
USE SENSOR A B
softkey, either the SENSOR A CAL FACTOR LIST table or the SENSOR B CAL
FACTOR LIST table is displayed. In the second line, a name for identifying the
sensor is displayed; this name can be defined via the
SENSOR LABEL
softkey.
The individual frequency points of the list are detailed below, but they can only
be entered in ascending frequency order. Next to the number of the frequency
point, there are two editable columns.
FREQUENCY – frequency value of the frequency point
CAL FACTOR – associated calibration factor in percentage
Linear interpolation occurs between the frequency points of the active
calibration factor list. The calibration factors of the lower or upper cutoff
frequency apply outside the specified frequency range. In the presetting, the
table is empty and no correction has been made, i.e. the calibration factor is
generally assumed to be 100%. If the list contains only one frequency point, the
FREQUENCY field cannot be edited, and CAL FACTOR then applies to all
frequencies.
IEC/IEEE bus command:
SYST:COMM:RDEV:PMET:CFAC:ASEN 2GHZ,99PCT,4GHZ,98PCT
DEL ALL
POINTS
The
DEL ALL POINTS
softkey deletes all values from the
table with the frequency-dependent insertion loss of the
power splitter.
IEC/IEEE bus command:
-
DEL ACTIVE
POINT
The
DEL ACTIVE POINT
softkey deletes the marked value
from the table with the frequency-dependent insertion loss of
the power splitter.
IEC/IEEE bus command:
-
INS NEW
POINT
The
INS NEW POINT
softkey inserts a new line where the
table is marked.
IEC/IEEE bus command:
-
GOTO
POINT #
The
GOTO POINT #
softkey sets the marking to the specified
line within the table.
IEC/IEEE bus command:
-
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...