Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1017
User Manual 1178.8520.02 ─ 01
14.8.3.10
Phase Noise Measurement Marker
The following commands control the phase noise measurement marker function.
Useful commands for phase noise markers described elsewhere
●
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:
PEAK]
●
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
●
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y
Remote commands exclusive to phase noise markers
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO
............................................. 1017
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
.......................................... 1017
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe]
........................................... 1018
CALCulate<n>:MARKer<m>:FUNCtion:PNOise:AOFF
.....................................................1018
CALCulate<n>:MARKer<m>:FUNCtion:PNOise:RESult?
................................................. 1019
CALCulate<n>:MARKer<m>:FUNCtion:PNOise[:STATe]
.................................................. 1019
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO
<State>
This command turns an automatic peak search for the fixed reference marker at the
end of a sweep on and off.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT:FUNC:PNO:AUTO ON
Activates an automatic peak search for the reference marker in
a phase-noise measurement.
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
This command queries the result of a phase noise measurement.
If necessary, the command activates the measurement first.
This command is only available in the Spectrum application.
Suffix:
<n>
.