Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1101
User Manual 1178.8520.02 ─ 01
String
Unit
'DBMV'
dBmV
'DBUV'
dB
μ
V
'DBUA'
dB
μ
A
'DBPW'
dBpW
'DBPT'
dBpT
MMEMory:LOAD<n>:TFACtor
<FileName>
Loads the transducer factor from the selected file in
.CSV
format.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
String containing the path and name of the
CSV
import file.
Example:
MMEM:LOAD:TFAC 'C:\TEST.CSV'
Manual operation:
See
MMEMory:STORe<n>:TFACtor
<FileName>, <TransdName>
This command exports transducer factor data to an ASCII (CSV) file.
For details on the file format see
Chapter 12.5.3, "Reference: Transducer Factor File
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
Name of the transducer factor to be exported.
<TransdName>
Name of the transducer factor to be exported.
Example:
MMEM:STOR:TFAC 'C:\TEST', 'Transducer1'
Stores the transducer factor named "Transducer1" in the file
TEST.CSV
.
Manual operation:
See
14.10.4
Compensating for Frequency Response Using Touchstone files
(R&S
FSV/A-K544)
If the Frequency Response Correction option (R&S
FSV/A-K544) is installed, the
R&S
FSV/A supports the use of Touchstone (
.SnP
) files, as well as additional fre-
quency response correction files in
.fres
format.
Configuring the R&S
FSV/A