Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1108
User Manual 1178.8520.02 ─ 01
[SENSe:]CORRection:FRESponse:USER:IQ:DATA:FREQuency?
[SENSe:]CORRection:FRESponse:USER:IQ:DATA:MAGNitude?
[SENSe:]CORRection:FRESponse:USER:IQ:DATA:PHASe?
Queries the trace values for the combined user correction files (
.snp
+
.fres
) in IQ
mode.
Example:
SENS:CORR:FRES:USER:IQ:DATA:PHAS?
Usage:
Query only
Manual operation:
See
[SENSe:]CORRection:FRESponse:BASeband:USER:LOAD
<FilePath>
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:LOAD
<FilePath>
[SENSe:]CORRection:FRESponse:USER:LOAD
<FilePath>
Loads a stored user-defined frequency response correction scenario.
Setting parameters:
<FilePath>
string
Example:
SENS:CORR:FRES:USER:LOAD 'FRes1'
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:FRESponse:BASeband:USER:PRESet
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:PRESet
[SENSe:]CORRection:FRESponse:USER:PRESet
Restores the default frequency response correction settings (containing only files spe-
cific to the R&S
FSV/A itself). Frequency response correction using
.fres
files is
deactivated.
Example:
SENS:CORR:FRES:USER:PRES
Usage:
Event
Manual operation:
See
[SENSe:]CORRection:FRESponse:USER:SCOPe
<Scope>
Determines whether the frequency response correction settings are applied to all
active measurement channels, or only the currently selected channel.
Parameters:
<Scope>
CHANnel | ALL
CHANnel
The frequency response correction settings are applied to the
currently selected channel only.
To select a channel, use
.
For a list of available channels, use
.
Configuring the R&S
FSV/A