Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1135
User Manual 1178.8520.02 ─ 01
OFF | 0
Disconnects the generator.
ON | 1
Connects the generator.
*RST:
0
Example:
CONF:GEN:IPC:ADDR '192.168.114.90'
CONF:GEN:CONN:STAT ON
Manual operation:
See
See
CONFigure:GENerator:IPConnection:ADDRess
<IPAddress>
The TCPIP address or computer name of the signal generator connected to the
R&S
FSV/A via LAN.
The IP address / computer name is maintained after a [PRESET], and is transferred
between applications.
Parameters:
<IPAddress>
IP address or computer name
Example:
CONF:GEN:IPC:ADDR '192.168.114.90'
Manual operation:
See
"IP Address or Computer name of Signal Generator"
14.10.10
Using Service Functions
....................................................................................1135
DIAGnostic:SERVice:SFUNction:LASTresult?
................................................................. 1136
DIAGnostic:SERVice:SFUNction:RESults:DELete
............................................................1136
DIAGnostic:SERVice:SFUNction:RESults:SAVE
..............................................................1136
.........................................................................................1136
......................................................................................1137
........................................................................................... 1137
DIAGnostic:SERVice:SFUNction
<ServiceFunction>
This command starts a service function.
The service functions are available after you have entered the level 1 or level 2 system
password.
Parameters:
<ServiceFunction>
String containing the ID of the service function.
The ID of the service function is made up out of five numbers,
separated by a point.
•
function group number
•
board number
•
function number
Configuring the R&S