Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1158
User Manual 1178.8520.02 ─ 01
.............................................................................................. 1159
.......................................................................................... 1160
SYSTem:HPCoupling
<CouplingType>
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW) and
●
resolution bandwidth and video bandwidth (RBW/VBW)
This command is only available if a HP language is selected using
Parameters:
<CouplingType>
HP | FSP
*RST:
FSP
Example:
SYSTem:HPC HP
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
Parameters:
<Mode>
NORM | PULS
NORM
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULS
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Manual operation:
See
SYSTem:HPADditional
<State>
Allows the use of HP commands
in addition to
SCPI commands for R&S
Parameters:
<State>
ON | OFF | 1 | 0
OFF | 0
Switches the function off
Emulating Other Instruments' Commands