Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1186
User Manual 1178.8520.02 ─ 01
14.13.2.3
Special Behavior of Commands
Command
Known Differences
ABORT
Does not automatically set the command complete bit (bit 4) in the status byte. An addi-
tional DONE is required for that purpose.
ANNOT
Only frequency axis annotation is affected.
AT
AT DN/UP: Step size
CAL
The CAL commands do not automatically set the command complete bit (bit 4) in the sta-
tus byte. An additional DONE command is required for that purpose.
CF
Default value, range, step size
CR
Default ratio Span/RBW
CT
Formula for coupled sweep time
CV
Default ratio RBW/VBW
DEMOD
requires option R&S
FSV3-B3
DET
DET? returns SAMP instead of SMP on the R&S
FSV/A.
DET not automatically set the command complete bit (bit 4) in the status byte. An addi-
tional DONE is required for that purpose.
ERR?
Deletes the error bit in the status register but always returns a '0' in response.
FA
Default value, range, step size
FB
Default value, range, step size
ID
Query of instrument ID. The instrument ID defined in "SETUP > N Remote >
GPIB > Identification String" is returned.
M2
Default value, range, step size
M3
Default value, range, step size
MKACT
Only marker 1 is supported as the active marker.
MKBW
Default value
MKPT
Step size
MKPX
Step size
OL?
Storage of instrument settings:
80 characters are returned as information on the instrument settings.
The contents of the 80 characters returned does not correspond to the original data con-
tents of the 8566A/8568A family.
OL
Readout of instrument settings:
The 80 characters read by means of OL? are accepted as information on the correspond-
ing instrument settings.
The contents of the 80 characters read does not correspond to the original data contents
of the 8566A/8568A family.
RB
Default value, range, step size
RL
Default value, step size
Emulating Other Instruments' Commands