Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
172
User Manual 1178.8520.02 ─ 01
If disabled, the upper and lower channels can be configured differently. This is particu-
larly useful if the lowest Tx channel and highest Tx channel use different standards and
thus require different bandwidths for adjacent channel power measurement.
Remote command:
Limit Checking
Activates or deactivates limit checks globally for all adjacent and gap channels. In addi-
tion to this setting, limits must be defined and activated individually for each channel.
The results of the power limit checks are also indicated in the
STAT:QUES:ACPL
sta-
tus registry (see
"STATus:QUEStionable:ACPLimit Register"
Remote command:
CALCulate<n>:LIMit<li>:ACPower[:STATe]
7.3.5.2
MSR Sub Block and Tx Channel Definition
Access
: "Overview" > "Select Measurement" > "Channel Power ACLR" > "CP / ACLR
Standard" > "Standard" : "Multi-Standard Radio" > "CP / ACLR Config" > "Tx Chan-
nels" tab
The "Tx Channels" tab provides all the channel settings to configure sub blocks and Tx
channels in MSR ACLR measurements.
Channel Power and Adjacent-Channel Power (ACLR) Measurement