Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
204
User Manual 1178.8520.02 ─ 01
For details on limit lines for searches, see
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
Deactivating All Search Limits
Deactivates the search range limits.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
7.5.4
How to Determine the Occupied Bandwidth
The following step-by-step instructions demonstrate how to determine the occupied
bandwidth.
For remote operation, see
Chapter 14.5.5.2, "Programming Example: OBW Measure-
How to determine the OBW for a single signal
1. Press the [MEAS] key or select "Select Measurement" in the "Overview" .
2. Select the "OBW" measurement function.
The measurement is started immediately with the default settings.
3. Select the "OBW Config" softkey.
The "Occupied Bandwidth" configuration dialog box is displayed.
4. Define the percentage of power ( "% Power Bandwidth" ) that defines the band-
width to be determined.
5. If necessary, change the channel bandwidth for the transmission channel.
6. To optimize the settings for the selected channel configuration, select "Adjust Set-
tings" .
7. Start a sweep.
The result is displayed as OBW in the marker results.
How to determine the OBW for a multicarrier signal using search limits
1. Press the [MEAS] key or select "Select Measurement" in the "Overview" .
2. Select the "OBW" measurement function.
3. Select the "OBW Config" softkey.
Occupied Bandwidth Measurement (OBW)