Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
243
User Manual 1178.8520.02 ─ 01
Signal generator settings (e.g. R&S
FSV/A SMW):
Device
Standard
Center frequency
Level
Test model
SigGen 1
3GPP/FDD
900 MHz
0 dBm
1-16
SigGen 2
EUTRA/LTE
906.5 MHz
0 dBm
1_1_5MHz
Setting up the measurement
1. Preset the R&S
FSV/A.
2. Set the center frequency to
903.25 MHz
.
3. Set the reference level to
10 dBm
with an offset of to
30 dB
.
4. Press the [MEAS] key or select "Select Measurement" in the "Overview" .
5. Select the "SEM" measurement function.
6. Select the "Sub Blocks" softkey and enter "Sub Block Count" of
2
.
7. For "Sub Block A" , define the settings for the 3GPP/FDD signal:
● Set the "Center Frequency" to
900 MHz
● Select "MSR Settings" .
● Set the "Base Station RF Bandwidth" to
5 MHz
.
● Select "Apply to SEM" .
8. For "Sub Block B" , define the settings for the EUTRA/LTE signal:
● Set the "Center Frequency" to
906.5 MHz
● Select "MSR Settings" .
● Set the "Base Station RF Bandwidth" to
5 MHz
.
● Select "Apply to SEM" .
9. Select [RUN SINGLE] to perform a measurement with the new settings.
The results of the measurement for each sub block are displayed in the Result
Summary. The summarized limit line is indicated in the Spectrum graph.