Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
244
User Manual 1178.8520.02 ─ 01
Figure 7-40: Multi-SEM measurement: results of the measurement for each sub block
7.6.8
Reference: SEM File Descriptions
This reference provides details on the format of the SEM settings and result files.
Format Description of SEM XML Files
..................................................................244
ASCII File Export Format (Spectrum Emission Mask)
.......................................... 250
7.6.8.1
Format Description of SEM XML Files
The SEM XML files offer a quick way to change the measurement settings. A set of
predefined XML files for different standards is already provided. You can also create
and use your own XML files. Alternatively, edit the settings directly in the "Spectrum
Emission Mask" dialog box and save the XML file afterwards. This way, you do not
have to modify the XML file itself.
In addition to saving the current settings to a file, settings files can also be created
independently of the R&S
FSV/A, in an external application. When creating your own
XML files, be sure to comply with the following conventions because the R&S
FSV/A
can only interpret XML files of a known structure. For sample files, see the
C:\Program Files\Rohde-Schwarz\FSV3000\<version>\sem_std
directory
of the R&S
FSV/A.
To load a settings file, use the "Load" function in the "Standard Files" tab of the "Spec-
trum Emission Mask" dialog box (see
"How to load an SEM settings file"
All XML files are stored under
C:\Program Files\Rohde-Schwarz\FSV3000\<version>\sem_std
.
The files for importing range settings obey the rules of the XML standard. The child
nodes, attributes, and structure defined for the data import are described here.
Spectrum Emission Mask (SEM) Measurement