Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
248
User Manual 1178.8520.02 ─ 01
Table 7-14: Attributes and child nodes of the PowerClass element
Child Node
Attribute
Value
Parameter description
Mand.
StartPower
Value
<power in dBm>
The start power must be equal
to the stop power of the previ-
ous power class. The Start-
Power value of the first range
is -200
Yes
Unit
dBm
Yes
InclusiveFlag
true
Yes
StopPower
Value
<power in dBm>
The stop power must be equal
to the start power of the next
power class. The StopPower
value of the last range is 200
Yes
Unit
dBm
InclusiveFlag
false
Yes
DefaultLimitFailMode
Absolute | Relative
| Absolute and Rel-
ative | Absolute or
Relative
Yes
Table 7-15: Attributes and child nodes of the Range element (normal ranges)
Child node
Attribute
Value
Parameter description
Mand.
Index
0…19
Indices are continuous and
have to start with 0
Yes
Name
<string>
Name of the range
Only if Referen-
ceChannel con-
tains a name and
the range is the
reference range
Short-
Name
<string>
Short name of the range
No
ChannelType
TX | Adjacent
Yes
WeightingFilter
Only if Referen-
cePower method
is TX Channel
Power and the
range is the ref-
erence range
Type
RRC | CFilter
Type of the weighting filter
Yes
Roll Off Factor
0…1
Excess bandwidth of the fil-
ter
Only if the filter
type is RRC
Bandwidth
<bandwidth in Hz>
Filter bandwidth
Only if the filter
type is RRC
FrequencyRange
Yes
Start
<frequency in Hz>
Start value of the range
Yes
Stop
<frequency in Hz>
Stop value of the range
Yes
Spectrum Emission Mask (SEM) Measurement