Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
271
User Manual 1178.8520.02 ─ 01
Edit Gate Ranges
Opens a dialog box to configure up to 3 gate ranges for each trace. For details see
Chapter 7.8.5.2, "Gate Range Definition for APD and CCDF"
Adjust Settings
Adjusts the level settings according to the measured difference between peak and min-
imum power for APD measurement or peak and mean power for CCDF measurement
in order to obtain maximum power resolution. Adjusts the reference level to the current
input signal.
Remote command:
CALCulate<n>:STATistics:SCALe:AUTO ONCE
7.8.5.2
Gate Range Definition for APD and CCDF
Access
: "Overview" > "Select Measurement" > "APD" / "CCDF" > "APD Config" /
"CCDF Config" > "Edit Gate Ranges"
You can configure gate ranges for gated triggering in statistical measurements.
For background information on defining gate ranges see
CCDF Basics - Gated Triggering"
The remote commands required to perform these tasks are described in
ter 14.5.8.3, "Using Gate Ranges for Statistical Measurements"
Up to three ranges can be defined for each of the six available traces.
.................................................................................................272
Comment
An optional comment can be defined for the gate range settings of each trace.
Statistical Measurements (APD, CCDF)